Search Results
Found 1 results
510(k) Data Aggregation
(21 days)
The MD 700 is a multi-parameter, automated hematology analyzer used to perform in vitro analysis of human peripheral blood in clinical laboratories.
The MD 700 reports on the following parameters:
White Blood Cells (Leukocytes) WBC
Red Blood Cells (Erythrocytes) RBC
Hemoglobin Concentration Hgb
Hematocrit (relative volume of erythrocytes) Hct
Mean Corpuscular (erythrocyte) Volume MCV
Mean Corpuscular (erythrocyte) Hemoglobin MCH
Mean Corpuscular (erythrocyte) Hemoglobin Concentration MCHC
Red Blood Cell (erythrocyte volume) Distribution Width RDW
Platelet or Thrombocyte Count Plt
Granulocyte (number) GR#
Granulocyte (percent of WBC) GR%
Lymphocyte (number) LY#
Lymphocyte (percent of WBC) LY%
Monocyte (number) MO#
Monocyte (percent of WBC) MO%
The MD 700 is a quantitative analyzer of peripheral human blood for in vitro diagnostic use in clinical laboratories. It performs an automated complete blood count (CBC) and a leukocyte differential on peripheral human blood. A sample volume of 20 uL of whole blood is required. The instrument provides a printed report on 15 parameters. The MD 700 consists of five components: (1) A blood analysis instrument, (2) A reagent dispensing unit and four reagents, (3) Calibration and control reagents, (4) Probeclenz™, a cleaning liquid, and (5) A standard, off-the-shelf high resolution printer. The MD 700 analyzes human peripheral blood by passing electronic waves through precisely measured dilutions of the blood, and analyzing the results using software which relies upon a patented mathematical algorithm to determine the count and percent of each parameter reported.
The provided document, K983562, is a 510(k) Summary for the MASCOT MD Hematology Analyzer Model MD 700. It establishes substantial equivalence to a predicate device (MASCOT MD Hematology Analyzer Model MD 800) and describes the device's technical specifications and intended use. However, it does not contain specific acceptance criteria, performance study details, or data proving the device meets acceptance criteria.
Instead of a new study, the document explicitly states: "This is a special 510(k). The non-clinical and clinical studies upon which the 510(k) of the MD 800 and its reagents were cleared are relied upon with respect to the MD 700. The MD 700 is as safe and effective as the MD 800, the predicate device, because the fundamental scientific technology of the MD 700 has not changed from that of the MD 800."
Therefore, I cannot provide a table of acceptance criteria and reported device performance or information about a specific study for the MD 700 directly from this document. The document relies on the prior approval of the predicate device (MD 800) for its safety and effectiveness claims.
I can, however, extract the performance specifications of the MD 700 as stated in the document, which can be inferred as a form of acceptance criteria for the device's operational characteristics.
Here's a breakdown of the requested information based on the provided text, noting where specific details are not present:
1. Table of Acceptance Criteria and Reported Device Performance
As explained above, the document does not explicitly state "acceptance criteria" as a separate set of performance targets. Instead, it provides "Significant Performance Characteristics of Device" (Page 5) and "MASCOT ™ MD 700 HEMATOLOGY SYSTEM SPECIFICATIONS" (Page 6), which serve as the reported performance of the device and implicitly as the performance benchmarks it is expected to meet.
| Characteristic | Acceptance Criteria (from document, implicitly) | Reported Device Performance (from document) |
|---|---|---|
| Precision | ||
| WBC | <3.0% c.v. | <3.0% c.v. @ 7.0 - 10.0 x 10³/μL |
| RBC | <3.0% c.v. | <3.0% c.v. @ 4.0 - 5.0 x 10⁶/μL |
| Hb | <2.0% c.v. | <2.0% c.v. @ 12.0 - 15.0 g/dL |
| MCV | <1.0% c.v. | <1.0% c.v. @ 80.0 - 90.0 fL |
| PLT | <6.0% c.v. | <6.0% c.v. @ 200 - 400 x 10³/μL |
| Linearity | ||
| WBC | 0.1 - 200.0 x 10³/μL, Limits: 0.4 or 5.0% | 0.1 - 200.0 x 10³/μL, Limits: 0.4 or 5.0% |
| RBC | 0.01 - 18.00 x 10⁶/μL, Limits: 0.10 or 8.0% | 0.01 - 18.00 x 10⁶/μL, Limits: 0.10 or 8.0% |
| Hb | 0.1 - 26.0 g/dL, Limits: 0.3 or 3.0% | 0.1 - 26.0 g/dL, Limits: 0.3 or 3.0% |
| MCV | 40.0 - 290.0 fL, Limits: 1.5 or 5.0% | 40.0 - 290.0 fL, Limits: 1.5 or 5.0% |
| PLT | 1 - 2000 x 10³/μL, Limits: 15 or 10% | 1 - 2000 x 10³/μL, Limits: 15 or 10% |
| Sample Volume | 20μL | 20μL |
| Throughput | 12 Specimens per Hour | 12 Specimens per Hour |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document states: "The non-clinical and clinical studies upon which the 510(k) of the MD 800 and its reagents were cleared are relied upon with respect to the MD 700." This means specific details about sample size, country of origin, or retrospective/prospective nature of the test set for the MD 700 are not provided. The data provenance and sample sizes would be found in the 510(k) submission for the predicate device, K981568 (MD 800), which is not included here.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable, as a new study for the MD 700, including establishing ground truth for a test set, was not conducted or reported in this document. It relies on the predicate device's clearance.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as a new study for the MD 700, including an adjudication method for a test set, was not conducted or reported in this document. It relies on the predicate device's clearance.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The MASCOT MD 700 is an automated hematology analyzer, not an AI-assisted diagnostic device requiring human reader interpretation in the context of an MRMC study. It provides automated counts and percentages of blood parameters.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, the device is described as performing an "automated complete blood count (CBC) and a leukocyte differential on peripheral human blood" (Page 1). The description of "How the MD 700 functions" (Page 2-4) clearly indicates a standalone operation where "the instrument automatically performs all remaining functions necessary to complete the analysis of the blood sample. The operator has no other function except appropriate handling of the printed report." The leukocyte differential counts are obtained from volume and intracellular complexity measurements and "processed according to a patented mathematical algorithm termed Expectation Maximization" (Page 5). This confirms standalone (algorithm-only) performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not specified for the MD 700. The device relies on the studies conducted for the predicate MD 800. Automated hematology analyzers typically validate their measurements against reference methods, which could involve manual microscopy (expert consensus) or other validated automated systems. However, this specific document does not detail the ground truth method.
8. The sample size for the training set
Not applicable for a new study on the MD 700. The document indicates the device uses a "patented mathematical algorithm named Expectation Maximization (EM)" (Page 3). If this algorithm involved machine learning that required a training set, the details of that training would be related to the development of the algorithm itself, not a specific "training set" for the MD 700's performance validation in this 510(k). The document relies on the predicate for validation.
9. How the ground truth for the training set was established
Not applicable, as detailed above. Details regarding "ground truth for the training set" would pertain to the development of the core algorithm, which is relied upon from the predicate device and not further elaborated in this 510(k) submission for the MD 700.
Ask a specific question about this device
Page 1 of 1